Aortic valve replacement with the medtronic mosaic bioprosthesis: a 13-year follow-up

Abstract

This study evaluated the long-term clinical performance of the Mosaic bioprosthesis (Medtronic Inc, Minneapolis, MN) after aortic valve replacement.From 1995 to 2008, 178 patients (48 women; mean age, 74±6 years) had aortic valve replacement. Mean functional class was 2.3±0.5, and 157 patients (88\%) were in sinus rhythm. Prosthetic sizes were 23 mm in 98 patients and 25 mm in 66. Follow-up was completed in December 2009 with a cumulative duration of 1,015 patient/years (mean, 5.7±3.5 years, maximum, 13.7 years).Early mortality was 4\%, none being valve-related; of 38 late deaths 7 were valve-related. Actuarial survival at 13 years was 48\%±8\%. Mean functional class of current survivors was 1.2±0.6. Six embolic episodes occurred and four cases of endocarditis, with respective actuarial freedom of 92\%±5\% for embolism and 97\%±2\% for endocarditis at 13 years. Four patients required reoperations for endocarditis and 2 for structural deterioration. Actuarial freedom from structural deterioration and from reoperation for all causes was 89\%±7\% and 86\%±7\% at 13 years, with an actuarial freedom from prosthesis-related deaths of 86\%±5\%. Results of echocardiographic evaluation at 1 year were mean peak gradient, 20±6 mm Hg and mean effective orifice area index, 1.07±0.21 cm2/m2 for size 23 mm and 22±6 mm Hg and 1.11±0.26 cm2/m2 for size 25 mm; at 10 years, mean peak gradient and mean effective orifice area index were 28±13 mm Hg and 1.01±0.19 cm2/m2 for size 23 mm and 26±8 mm Hg and 1.08±0.18 cm2/m2 for size 25 mm.The Mosaic bioprosthesis showed good overall performance, with low incidence of structural valve deterioration and hemodynamic stability in the long-term. Expected increased durability of this device should be verified at longer follow-up intervals.


Autore Pugliese

Tutti gli autori

  • MILANO A.D.

Titolo volume/Rivista

Non Disponibile


Anno di pubblicazione

2012

ISSN

0003-4975

ISBN

Non Disponibile


Numero di citazioni Wos

Nessuna citazione

Ultimo Aggiornamento Citazioni

Non Disponibile


Numero di citazioni Scopus

20

Ultimo Aggiornamento Citazioni

Non Disponibile


Settori ERC

Non Disponibile

Codici ASJC

Non Disponibile