A biocompatible cardiopulmonary bypass strategy to reduce hemostatic and inflammatory alterations: a randomized controlled trial.
Abstract
Cardiopulmonary bypass (CPB) systems without a venous reservoir rarely are adopted clinically. The effects of a biocompatible CPB system with a venous reservoir were evaluated on the activation of the coagulation and inflammatory systems. DESIGN: A prospective, randomized controlled trial. SETTING: A university hospital (single center). PARTICIPANTS: Eighty-three coronary artery bypass graft (CABG) surgery patients were assigned to the Physio group (closed venous reservoir, phosphorylcholine coating, and no cardiotomy suction) or the Standard group (open, noncoated, and cardiotomy suction used). METHODS: Blood samples were obtained at 6 different time points before, during, and after surgery. Nuclear factor-kB (NF-κB) was evaluated before surgery and 2 and 24 hours after surgery. Myocardial damage was evaluated measuring cardiac troponin I. MEASUREMENTS AND MAIN RESULTS: Interleukin (IL)-6 (a marker of inflammation), prothrombin fragment 1-2 (PF-1.2, a marker of thrombin generation), plasmin-antiplasmin complex (PAP, a marker of fibrinolysis), and platelet factor 4 (PF4, a marker of platelet activation) were measured. The DNA binding activity of proinflammatory transcription factor NF-κB was quantified in the isolated lymphomonocyte cells. Surgery caused changes of all plasma biomarkers. This reaction was attenuated strongly in the Physio group; PF-1.2, PAP, and PF4 all were decreased significantly. In the Physio group, a significantly lower cardiac troponin I release was observed postoperatively. After surgery, NF-κB activity was reduced in the Physio group although this difference was not statistically significant. CONCLUSIONS: A multimodal strategy using a closed and phosphorylcholine-coated CPB circuit together with the avoidance of cardiotomy suction reduced activation of the coagulation and fibrinolytic systems intraoperatively, although these changes did not persist postoperatively. However, no difference in clinical outcome was appreciated on a larger scale.
Autore Pugliese
Tutti gli autori
-
PAPARELLA D.
Titolo volume/Rivista
Non Disponibile
Anno di pubblicazione
2012
ISSN
1053-0770
ISBN
Non Disponibile
Numero di citazioni Wos
Nessuna citazione
Ultimo Aggiornamento Citazioni
Non Disponibile
Numero di citazioni Scopus
Non Disponibile
Ultimo Aggiornamento Citazioni
Non Disponibile
Settori ERC
Non Disponibile
Codici ASJC
Non Disponibile
Condividi questo sito sui social